EasyManuals Logo

Sartorius Mark 3 User Manual

Sartorius Mark 3
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
Serial Command List
Command Format Limit Description Response Format Range
Test Setting Commands
SET UNITS MOISTURE UNITS MOISTURE
SOLIDS UNITS SOLIDS
VOLATILES UNITS VOLATILES
MOISTURE/DRY UNITS MOISTURE/DRY
PPM MOISTURE* UNITS PPM MOISTURE
PPM SOLIDS* UNITS PPM SOLIDS
WEIGHT UNITS WEIGHT
SET TEMP1 + Float (0-210) 0=OFF TEMP1 x 25-210C
SET TEMP2 + Float (0-210) 0=OFF TEMP2 x 25-210C
SET RAMP1 STANDARD* “STANDARD”/”RAPID” RAMP1 STANDARD
RAPID* RAMP1 RAPID
SET RAMP2 STANDARD* “STANDARD”/”RAPID” RAMP2 STANDARD
RAPID* RAMP2 RAPID
SET TIME1 xx�x Float (0-99�9) 0=OFF TIME1 xx�x 0=99�9 minutes, 0=OFF
SET TIME2 xx�x Float (0-99�9) 0=OFF TIME2 xx�x 0=99�9 minutes, 0=OFF
SET FACTOR x�xxx Float FACTOR x�xxx
SET OFFSET x�xxx Float OFFSET x�xxx
SET PLACES x Integer (2 or 3) PLACES x 2,3
SET SLOPE x�xxx Float (0�0 - 9�99) SLOPE x�xxx 0-9�99%, 0=OFF
SET SLOPETIME x�xx Float (0�1 - 99�99) SLOPETIME x�xx 0-9�99 minutes, 0=OFF
SET CALC ACTUAL “ACTUAL”/”CALC1”/
”CALC2”/”CALC3” CALC ACTUAL
CALC1 CALC CALC1
CALC2 CALC CALC2
CALC3 CALC CALC3
SET DELAY x Integer (0 - 9) DELAY x 0-9 sec
SET STANDBY x Integer (<165) STANDBY x 25-165C
SET EQUILIBRATE x* Integer (0 - 60) EQUILIBRATE x 0-60 minutes
SET IDEALWEIGHT xx�x Float (0�1 - 99�0) IDEALWEIGHT xx�x 0-99 grams
SET IDEAL PERCENT xx�x Float (0�1 - 10�0) IDEAL PERCENT xx�x 1-10
SET IDEALLOCK ON “ON”/”OFF” IDEALLOCK ON
OFF IDEALLOCK OFF
SET TARGETUPPER xxx�x Float (0 - 100�0) TARGETUPPER xxx�x 0-100
SET TARGETLOWER xxx�x Float (0 - 100�0) TARGETLOWER xxx�x 0-100
SET TARGETLIMIT ON “ON”/”OFF” TARGETLIMIT ON
OFF TARGETLIMIT OFF
SET MODE STANDARD “STANDARD”/”SYRINGE”/
”CONCENTRATION” MODE STANDARD
SYRINGE* MODE SYRINGE
CONCENTRATION MODE CONCENTRATION
SET PROGRAMNAME string String PROGRAMNAME string Max: 15 characters
GET TEMP1 Float (0-210) 0=OFF TEMP1 x
GET TEMP2 Float (0-210) 0=OFF TEMP2 x
* = only for model LMA100P
54