Hazards for persons or equipment 
posed by using specific samples:
 
 
 Fire  Explosion
–  Flammable or explosive substances
–  Substances that contain solvents
–  Substances that release flammable or 
explosive gases or vapors during the 
drying process
  In some cases, it is possible to operate 
the moisture analyzer in an enclosed 
nitrogen atmosphere to prevent the 
vapor released during drying from  
coming in contact with oxygen in the 
surrounding atmosphere� Check on a 
case-to-case basis whether this method 
can be used, because installation of the 
analyzer in too small an enclosed space 
can affect its functions (for instance 
through excessive heat build-up within 
the analyzer)� When in doubt, perform  
a risk analysis�
  The user shall be liable and responsible 
for any damage that arises in connec-
tion with this moisture analyzer�
 
 
  Poisoning  Caustic burns
–  Substances containing toxic or caustic 
or corrosive substances: These may  
be dried only under a fume hood�  
The value for the “lower toxic limit” in  
a work area must not be exceeded� 
Corrosion:
–  Substances that release aggressive 
vapors during the heating process  
(such as acids): In this case we recom-
mend that you work with small sample 
quantities� Otherwise, vapors can con-
dense on cold housing parts and cause 
corrosion�
  The user shall be liable and responsible 
for any damage that arises in connec-
tion with this moisture analyzer�
–  Sample Pans (Delivered with the 
 Analyzer): 
The reusable metal pans are not  
made of stainless steel� For this reason, 
they are not resistant to corrosion� 
These panscan be used only for pow-
der and granules (e�g� plastics) with very 
low moisture content�
  If the moisture content is higher, the 
disposable aluminium pans (90 mm 
diameter) must be used (see chapter 
“Accessories”)�
6