61
Conserve Mode
  The Conserve Mode allows the analyzer to manage the 
heating elements to be turned Off when the analyzer 
is not in use� In addition, a time schedule can be set 
for each day of the week for a Sleep mode and a Wake 
mode�
  To edit or change the Conserve Mode period:
§  From the Setup Menu screen, press the 7 key or high-
light the Conserve mode using the navigation keys and 
press Enter to display the Conserve Mode screen�
§  Press the 1 key for Conserve to show the Conserve 
screen�
§  Use the numeric keys to type in a period of time up to 
60 minutes and press Enter to return to the Conserve 
Mode screen�
Note
  To turn Conserve Off, meaning the analyzer will main-
tain the standby temperature while the analyzer is not 
being used, enter 0 minutes�
  To set or change the sleep and wake mode:
§  From the Conserve Mode screen press 2 or 3 key to 
show the Sleep mode or Wake up mode screen�
§  From the Sleep mode or Wake up mode screen, type in 
the desired time (use AM|PM key to set correctly) and 
press the Enter key�
§  Press the Day soft key� Then select the days of the week 
that the Sleep mode or Wake up mode is to be used� 
You will need to press the appropriate number key for 
each day to toggle it either On or Off�
§  Press the Enter key to return to the Wake up or Sleep 
mode screen�
§  Press the Enter again to return to the Conserve Mode 
screen�
  To manually wake the analyzer, press the Enter key�
1. Conserve         60  minutes 
2. Sleep  mode        Off 
3. Wake  up  mode    Off 
Conserve mode 
09/07/2012 02:14  PM 
On/Off