Default Drying Program
Your analyzer is delivered with a default drying pro-
gram� The drying program contains all the specific
parameters to test a sample including temperature,
endpoint, unit of measure, standby temperature,
etc� You will notice if you are in the default program or
a specific material program by the program name
at the top left of the display�
24
Program name: 01Default
Temp1=105C, Standard ramp,
Time1=Off
Temp2=Off, Time2=Off
Slope=0�050%/1 minute, Actual
Start delay=0 second
Standby temp=60C,
Equilibration=Off*
Ideal weight=5g, Lock=Off
Target limits=Off
Mode=Standard
* = only for model LMA100P