Functions available from the Standby screen:
§ Press Start|Stop key to initiate a test
§ Press Recall soft key to change the program to be 
used for testing a sample
§ Press User soft key to select an operator name  
(feature must be turned on in Setup)
§ Press Setup soft key to enter the Setup menu in 
order to make all parameter changes
§ Press Graph soft key to display a graph
§ Press Weigh soft key to use as a balance
 Standby-Temperature 
!
 with a New Program
When creating a new program, the standby tem-
perature setting from the previous drying program 
is used automatically by the analyzer�
Example:
You have created a new program for drying  
a temperature-sensitive sample at low temperature� 
The analyzer uses the standby temperature, e�g� 
165°C, from the previous program�
You must call up the standby temperature param-
eter and change the temperature value� Ifyou for-
get performing this setting, your analyzer is always 
too hot before starting the test�
23