Cleaning
Recommended care
For best performance and reliable 
usage, the analyzer requires periodic 
cleaning�
Danger
Before beginning the cleaning process:
§  The analyzer must be disconnected 
from the power source�
§  Allow the analyzer to cool down for  
30 minutes�
§  Take care during cleaning not to spill 
any liquid into the analyzer�
§  The I/O connections on the back of the 
analyzer may not come into contact 
with liquids�
§  Do not immerse the analyzer in water or 
any other cleaning solution�
Regular cleaning of the analyzer with a 
mild detergent and a soft cloth should 
include the following:
§  Clean the outer surface of the analyzer�
§  Carefully remove and clean the pan 
support and pan shield�
§  Keep all warning labels clean�
§  Wipe dust from the interior of the  
stainless steel heater housing�
§  Remove and clean the recessed area 
under the pan shield:
  –   Remove the pan support and pan 
shield�
  –   Remove the two screws holding the 
perforated plate�
  –   Be careful not to spill liquid or solid 
debris down the center hole where 
the balance pan receiver protrudes�
82