Leveling the Moisture Analyzer
  In order to function properly, the heater 
module should be horizontal� The heater 
module has a level vial� You can tell 
when the analyzer is level by checking 
whether the bubble is centered in the 
vial�
§  Check that all four feet are in the fully 
inserted position�
§  Observing the level vial, rotate one foot 
at a time to level�
  Adjust control module so that it does 
not rock�
Note
  The heater module must be re-leveled 
each time the analyzer is moved�
Keypad
Warm-up
  To enable the analyzer to function  
optimally, plug it in and leave it pow-
ered on for a minimum of 15 minutes� 
The analyzer is preprogrammed to alert 
the operator that the device is in the 
warm-up period� It is best to allow the 
analyzer to complete this warm-up  
period before testing samples; however, 
this may be overridden by pressing the 
Enter key�
14
Numeric 
keys
Soft keys
Navigational
keys