Test Mode
  This feature provides a few special test 
modes for specific applications:
§  Standard: normal test mode where a 
sample is placed directly onto the pan 
or pan with sample pads to take a direct 
weight of the initial weight�
  LMA100P:
§  Syringe: used for volatile liquid samples 
to prevent evaporation of the sample 
prior to obtaining an initial weight�  
In this mode, special test mode prompts 
will guide the operator to weigh a full 
syringe containing a sample and then 
again after the syringe content is placed 
onto the sample pad� The analyzer 
automatically calculates the initial 
weight as the difference between the 
full and empty syringe� See the chapter 
on Syringe Mode in this manual�
§  Concentrate: used for determining total 
suspended solids in low solid samples� 
The sample solids are first concentrated 
onto a filter pad placed in a vacuum 
apparatus� During the test mode, the 
operator will then be prompted to enter 
the number of milliliters of sample that 
was filtered� Results are displayed in 
mg/l� See details in the chapter on  
Concentration Mode in this manual�
  To change the test mode:
§  Press the 9 key from the Program Menu 
to display the Mode screen�
§  Press the number key for the mode of 
choice� The display will return to the 
Program Menu�