71
Calibration | Adjustment  Functions
Calibrate Balance
The balance of the analyzer should be calibrated and 
adjusted on installation and then periodically at some 
interval consistent with your quality program (one year 
is typical)�  
LMA100P: The balance may be calibrated and adjusted 
using the internal 100g calibration weight or a known 
external calibration weight�  
LMA110S: The balance may be calibrated manually 
with a Known external calibration weight�  
The analyzer also allows you to set a calibration interval 
that will remind you when calibration is due�
To access the balance calibration and other balance test 
features:
§  From the Service menu screen, press the 2 key for  
“Calibrate Balance�”
To calibrate using the internal weight only to the 
LMA100P:
§  Remove any object from the pan support
§  From the Calibrate balance screen, press the 1 key  
for Calibrate internal weight to display the Calibrate 
internal screen�
§  The balance will automatically re-zero�
§  The analyzer will automatically activate the internal 
calibration and adjustment mechanism�
Calibrate  internal 
Abort 
Performing  internal 
weight  calibratio n 
1) Calibrate  internal  weight* 
2) Calibrate external weight 
3) Schedule  calibration 
4) Reproducibility  tes t* 
Calibrate  balance 
09/07/2012  02:14  PM 
* = only for model LMA100P