Target Limits
The Target limit Menu provides a statistical process 
control (SPC) feature that can be set to identify when 
a tested sample is within or outside the desirable limits 
selected� If this feature is activated, a Pass or Fail will 
be displayed after each test, indicating the sample is 
within or outside the set limits; this will also appear on 
the printout�
To turn the Target Limits feature on and to enter limits:
§  Press the 8 key from the Program Menu to display 
the Target Limits screen� The current target limits are 
shown�
§  Press the On|Off soft key to toggle target limits: On
§  Press the 1 key to set the upper limit�
§  Press the 2 key to set the lower limit�
When all the parameters for target limits have been 
changed, press the Arrow key to accept the changes� 
The display will return to the Program Menu�
Note only for LMA100P:
  The limits are set in the current unit with the  
exception of ppm Moisture and ppm Solids�  
These are set in percent� Example: If you wish to set a 
lower limit of 50 ppm M, enter 0�005�  
(ppm / 10,000 = %)
40
1. Upper Limit 100.00%    Moisture 
2. Lower Limit 0.00%       Moisture 
Targ et  Limits  Menu 
Default  09/07/2012  02:14  PM 
On/off