EasyManua.ls Logo

Sartorius Mark 3 - Ideal Weight Percent Limit; Ideal Weight Lock

Sartorius Mark 3
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ideal Weight Percent Limit
The ideal weight limit is a tolerance
range for the ideal weight in percent
that the sample weight must be in�
Example: If the ideal weight is 10 grams,
and the target limit is 10%, then an
acceptable sample weight is between
9 and 11 grams�
To set or change the Ideal weight per-
cent limit:
§ Press the % soft key to display the Ideal
weight percent screen�
§ Use the numeric keys to type in the
desired time, press the Enter key� The
display will return to the Ideal Weight
screen�
Ideal Weight Lock
When activated, the ideal weight lock
prevents the use of any other sample
weight other than the one that is within
the target limits� If the lock is not acti-
vated, when in the test mode, the target
weight guide is still shown but any
sample weight can be used for the test�
To set or change ideal weight lock:
§ Press the Lock soft key to toggle On or
Off on the Ideal Weight screen�
39

Table of Contents

Related product manuals