Each heater module can run a test inde-
pendent from the other heater mod-
ules by recalling a program from the
program list and initiating a test� The
program list is common for all heater
modules� When a test is initiated, the
display will show test screens from the
single module mode, such as entering
an ID number and bar graph prompt to
add the ideal sample size�
To test a sample in multi-mode:
§ Press the navigational right or left
arrow key to move the double bars to
the desired heater module�
§ Check current program� To change
program, press Recall and select a new
program if desired�
§ Press the Start|Stop key to initiate a
test� (the screens and functionality are
the same as in the single module mode)�
§ After the initial weight is captured, the
display will return to the multi-mode
screen� The status line will show the
elapsed time of the test to indicate a
test is in progress�
§ At the end of the test, the status line
will display “Test Done” (or Pass or Fail
if target limits have been entered in this
program)� The printer will also print the
result at the end of the test (LMA100P
only)�
Note
Certain functions are not available in
the multi-module mode including:
Interval print and graphing� However,
in the multi-module mode, one can
test one or more samples and still have
access to service for another heater
module which is not testing a sample�
To perform service options on any
heater module such as:
Balance calibration
Temperature adjustment
LMA100P:
Balance repeatability test
§ From the multi-module screen, press
the navigational right or left arrow key
to move the double bars to activate
the desired heater module�
§ Press the Service soft key to display the
Service menu�
§ Press the number key of the desired
option and press the Enter key�
The display will show these screens
as in the single module mode�
LMA100P: Operation using 3 or 4
heating modules
The inter-module cables between
control module and heater module are
so short that only one heater module on
the left and right of the control module
can be connected� For this case, please,
order the optional 69TC0131 cable kit�