81
Each heater module can run a test inde-
pendent from the other heater mod-
ules by recalling a program from the 
program list and initiating a test� The 
program list is common for all heater 
modules� When a test is initiated, the 
display will show test screens from the 
single module mode, such as entering 
an ID number and bar graph prompt to 
add the ideal sample size�
To test a sample in multi-mode:
§  Press the navigational right or left 
arrow key to move the double bars to 
the desired heater module�
§  Check current program� To change  
program, press Recall and select a new 
program if desired�
§  Press the Start|Stop key to initiate a 
test� (the screens and functionality are 
the same as in the single module mode)�
§  After the initial weight is captured, the 
display will return to the multi-mode 
screen� The status line will show the 
elapsed time of the test to indicate a 
test is in progress�
§  At the end of the test, the status line 
will display “Test Done” (or Pass or Fail 
if target limits have been entered in this 
program)� The printer will also print the 
result at the end of the test (LMA100P 
only)�
Note
Certain functions are not available in 
the multi-module mode including:  
Interval print and graphing� However,  
in the multi-module mode, one can 
test one or more samples and still have 
access to service for another heater 
module which is not testing a sample�
To perform service options on any 
heater module such as:
  –  Balance calibration
  –  Temperature adjustment
 LMA100P:
  –  Balance repeatability test
§  From the multi-module screen, press 
the navigational right or left arrow key 
to move the double bars to activate  
the desired heater module�
§  Press the Service soft key to display the 
Service menu�
§  Press the number key of the desired 
option and press the Enter key�  
The display will show these screens  
as in the single module mode�
  LMA100P: Operation using 3 or 4 
heating modules
The inter-module cables between  
control module and heater module are 
so short that only one heater module on 
the left and right of the control module 
can be connected� For this case, please, 
order the optional 69TC0131 cable kit�