Mark 2 Program Download Format
Field Description
1  Program Number (#XXX)
2  Units (M,S,W,V,L)
3  Ideal Weight (XX�X, Negative value indicates ideal wt� loading locked 90-110%)
4  Temp 1 (XXX)
5  Temp 2 (XXX)
6  Time 1 (Seconds)
7  Time 2 (Seconds)
8  Standby Temp (XXX)
9  Slope Mode (0=Actual, 1=Calc1, 2=Calc2, 3=Calc3)
10  Slope Time (Tenths of minutes, example: 50=5�0 minutes)
11  Slope % (XX�XXX)
12  Start Delay of Initial Wt� Capture*
13  Program Name (19 characters)
Output Example (Factory Defaults):
# 1,M, 5�0,105,  0,    0,    0, 60,0,10, 0�050,65,FACTORY 1
# 2,M,30�0,110,  0,  120,    0,110,0,20, 0�010,65,FACTORY 2
# 3,S, 2�0,135,  0,    0,    0,130,0,10, 0�100,65,FACTORY 3
# 4,W, 5�0,105,130,  300,  300, 60,0, 0, 0�000,65,FACTORY 4
# 5,L, 2�0,135,  0,    0,    0,100,0,10, 0�050,65,FACTORY 5
/r/n
* Field 12 - Start Delay of Initial Wt� Capture Bit 0���Sample ID 0=Off   1=On
Sample ID Off    Sample ID On
Time(s)  Value  Hex  Time (s)  Value  Hex
0  0  0000 0000  0  1  0000 0001
1  16  0001 0000  1  17  0001 0001
2  32  0010 0000  2  33  0010 0001
3  48  0011 0000  3  49  0011 0001
4  64  0100 0000  4  65  0100 0001
5  80  0101 0000  5  81  0101 0001
6  96  0110 0000  6  97  0110 0001
7  112  0111 0000  7  113  0111 0001
8  128  1000 0000  8  129  1000 0001
9  144  1001 0000  9  145  1001 0001
60