4�  At the “Add sample to pan” prompt, open the heater 
hood and begin adding sample to the pan� A bar graph 
on the display will guide the operator to place the 
correct amount of sample as set by the ideal weight 
parameter of the drying program� Continue adding 
sample until the beeper sounds, indicating that the 
sample amount is within 10% of the ideal weight� The 
exact weight is also displayed on the screen�
5�  At the “Close heater hood” prompt, close the heater 
hood� The balance will capture the initial weight of the 
sample “Sampling” and then show that the test is in 
progress� The display will change to show  
current information including the elapsed time and the 
current percentage weight loss in the units as specified 
in the drying program�
Note
  To stop the test in progress, press the Start|Stop key�
  The printer will begin to print as the test begins� Print-
out information can be selected in Setup menu under 
Printer (LMA100P only)�
  At the end of the test the display will show “Test 
Done�” The result will stay on the screen as well as be 
printed� The analyzer will enter the Standby mode until 
the next test is initiated�
27
   
Dry  Milk  Powder  09/07/2012  02:14  PM 
Test  in  Progre ss 
0.015
% Moisture
Elapsed  time:  00.10 
Graph 
Press  STOP  key  to  stop  test
Temp 1 
Current 
105 
71 
Press  START  ke y  to  begin  test 
Dry  Milk  Powder  09/07/2012  02:14  PM 
Test Done 
3.471 % Moisture
Result time:    7.0  min 
ID:  126399.1 
Target:  Pass 
Recall  Setup 
Graph User 
Wa rning:  Sample   on  pa n 
 
Standby 
Current 
105 