Table: Program Settings
Screen Parameter Variables Explanation
Units Units Moisture Initial wt� – final wt� / Initial wt� + 100
Solids Final wt� / Initial wt� + 100
Volatiles Initial wt� – final wt� / Initial wt� + 100
Moisture/dry Initial wt� – final wt� / Final wt� + 100
ppm moisture* % Moisture + 10,000
ppm solids* % Solids + 10,000
mg/l Final wt� + 1000 / volume in mL
Weight Initial wt� – final wt�
TGA moisture* See appendix
TGA solids* See appendix
Result display Decimal place 0�01% or 0�001%
Factors Scale factor Factored result = (result + scale factor) + offset
Offset Factored result = (result + scale factor) + offset
Temperature 1 Temperature Maximum 210°C Set temperature or Off
Ramp rate* Standard Standard temperature ramp rate
Rapid More rapid ramp rate
Time 10�0 – 99�9 minutes Set time or Off
Temperature 2 Temperature Maximum 210°C Set temperature or Off
Ramp rate* Standard Standard temperature ramp rate
Rapid More rapid ramp rate
Time 20�0 – 99�9 minutes Set time or Off
Slope % Initial weight 0�010 – 9�990 Weight variable of endpoint criteria
Window of time 0�1 – 99�9 minutes Time variable of endpoint criteria
Calculation Actual slope Slope criteria met for endpoint
Calc 1 Calculated endpoint
Calc 2 Calculated endpoint
Calc 3 Calculated endpoint
Start delay Start delay 0 - 9 seconds Delay before starting test
Mode Mark 2 Simulates Mark 2 start delay
LMA100P | LMA110S Start delay
Standby temp� Standby temp� Maximum 165°C Set temperature between tests
Equilibration time* 1 – 60 minutes Delay after standby temperature met
Ideal weight Ideal weight 0�1 – 99�0 g Set ideal sample weight for test
Percent 1 – 10% Set tolerance of ideal weight
Ideal weight Lock On or Off Set ideal weight lock strict (On) or Off
Target limit Target limit Upper limit Set upper limit of result tolerance
Lower limit Set lower limit of result tolerance
Turn limits Off Turns target limit feature Off
Mode Mode Standard Standard test mode
Syringe* Initial weight taken by difference
Concentration Corrects a result for sample volume
* = only for model LMA100P
31