INTECONT® Tersus Belt weigher, Instruction Manual
Schenck Process Group
BV-H2464GB, 1234
- 113 -
15
1515
15
Events
EventsEvents
Events
15.1
15.115.1
15.1
Events Details
Events DetailsEvents Details
Events Details
15.1.1
15.1.115.1.1
15.1.1
Event Group:
Event Group:Event Group:
Event Group:
Calibration
CalibrationCalibration
Calibration
CA01 Event: Error LC Input
(P14.06)
The load cell cable is improperly connected, not connected or has been interrupted.
The load cell may be faulty.
Steps to take:
- check the cabling and the load cell
- check the load cell
Note:
The power supply must be switched off and on again if the scales still cannot be
switched on even after the error has been rectified.
CA02 Event: Tachometer Input
(P14.03)
The signals at the feeder drive's impulse speedometer input are outside of the speci-
fied range.
Possible causes:
- there is a fault in the speed impulse sensor cable (check the shielding and the wir-
ing).
- the sensor has been incorrectly calibrated
CA05 Event: Linearization Error
(P15.10)
Not a monotonically increasing curve.
A monotonically increasing compensation curve is monitored when:
- linearization is switched on,
- a parameter is altered
CA06 Event: Tare Correction
(P08.05)
This event is identified if the change in tare identified by the automatic zeroing sys-
tem exceeds the threshold value in P08.04 Zeroing Limit.
Possible causes:
- low quantities of bulk material on the belt
- error in the weighing mechanics
Steps:
- visual inspection of the belt and the weighing system.
- run the base tare adjustment program
- check weight measuring using test weights