Parameters
BV-H2464GB, 1234
- 72 -
INTECONT® Tersus Belt weigher, Instruction Manual
Schenck Process Group
P06.04 Event: Actual Flow Rate MAX
Default: W2
Event Group: HI Number: 02
Actual feed rate higher than the maximum value set in P06.03 Limit: Actual Flow Rate MAX
Possible causes:
- bulk material shooting out of the discharge chute
- disturbance forces from force shunts or pressures acting on the weighing equipment
- feed rate setpoint too large
P06.05 Limit: Load MIN
Default: 20 % Q
Min: -10 % Q Max: 200 % Q
Lower limit for belt load.
Reference: P04.01 Nominal Belt Load
P06.06 Event: Load MIN
Default: W2
Event Group: LO Number: 03
The belt load is less than the minimum value set in Limit: Load MIN
Possible causes:
- no bulk material
- disturbance forces from force shunts or pressures acting on the weighing equipment
P06.07 Limit: Load MAX
Default: 120 % Q
Min: -10 % Q Max: 200 % Q
Upper limit for the belt load
Reference: P04.01 Nominal Belt Load
P06.08 Event: Load MAX
Default: W2
Event Group: HI Number: 03
The belt load is greater than the maximum value set in P06.07 Limit: Load MAX
Possible causes:
- bulk material shooting out of the discharge chute
- disturbance forces from force shunts or pressures acting on the weighing equipment