5. INSTRUCTIONS FOR USE
14.
5. INSTRUCTIONS
FOR USE
TURNING ON THE
UNIT
Once the sterilizer has been correctly installed, it may be turned on and prepared for use.
Turn on the Bravo using the main (luminous) switch located on the right side of the unit.
INITIAL
AUTOMATIC
TEST
When turned on, the control panel lights up and beeps so you can visually check its correct
operation. The panel then displays the message:
BRAVO17
R. Exxxx/BMyyyyyy
DEVICE CHECK- UP
IF THE DOOR IS CLOSED, THE TEST IS INTERRUPTED. THE PANEL WILL BEEP AND
DISPLAY THE FOLLOWING MESSAGE.
OPEN DOOR
TO CONTINUE
Open the door to allow the test to continue. At the end of the test you will see:
BRAVO17
R. Exxxx/BMyyyyyy
CHECK- UP COMPLETE
AMBIENT
PRESSURE
VALUES
The sterilizer measures the ambient pressure for the correct operation of several auxiliary
devices. Whenever the difference between the value read and that previously stored is higher
than a set value, the system automatically updates the stored value after a brief delay.
Otherwise, the data remains unchanged without updating. (see Chapter 6 – Configuration -
Acquisition of the ambient pressure)
After updating, the device performs the initial automatic test procedure (see above). At the
end, the display shows the following message (accompanied by a beep):
AMBIENT PRESSURE
VALUE UPDATED
- 0.01 bar
↵ to continue
When ↵ is pressed, the device goes to STAND-BY mode (see below).