5. INSTRUCTIONS FOR USE
17.
MAX LEVEL IN THE
INTERNAL DRAIN
TANK
Emptying the internal
tank
When the water level in the internal drain tank reaches the MAX level, the LCD display
alternatively flashes the MAX and MIN icons.
NOTE
AT THIS STAGE, THE UNIT WILL GENERATE AN ALARM INDICATION (SEE APPENDIX E -
AL AR MS) SHOULD YOU ATTEMPT TO S T A R T A STERILIZATION CYCLE.
When this occurs, empty the internal and/or external draining tank.
To drain the internal tank, follow these steps :
1. Arrange an empty container on the floor near the
sterilizer and put the free end of the supplied tube
into the container.
2. Connect the quick connector to the corresponding
female connector under the chamber entrance
(marked ), pushing until you hear a click.
3. Wait for the internal tank to drain completely, then
while pinching the tube with your fingers, press the
metal lever located on the side of the connector
and detach the quick connector.