Analysis and resolution of problems
If your sterilizer is not working correctly, please consult this list before calling the Technical Service Department:
The sterilizer does not
power-on.
The power cord is not plugged-in.
There is no voltage at the socket.
Check the cause for the lack of voltage at the socket
and fix it.
The main switch and/or differential
switch are OFF.
The mains fuses are blown.
Replace with good fuses of equal nominal value.
(See the Summary Table in Appendix A, Technical
Characteristics).
After pressing S TA RT ,
the sterilization cycle
does not start.
The device is preheating.
Wait for the sterilizer to reach the proper operating
conditions for starting the program.
NOTE: Under normal conditions, the average
preheating time is approx. 10-15 minutes.
The MIN water level
icon is lit.
The distilled water level inside the
tank is below the minimum level.
Fill the distilled water tank until the MAX level indicator
comes on (or, at least, until the MI N level signal turns
off).
An alarm was triggered and an
error code and message appear on
the LCD.
Check the alarm code and take the appropriate action.
(See the following paragraphs, Alarms, Alarm Codes
and Troubleshooting).
The safety valve wa s
triggered.
Locking ring loosened.
Presence of anomalous
overpressure in the chamber.
Check that the knurled locking ring is correctly
tightened on the upper part of the safety valve.
LET THE DEVICE COOL, OR
WEAR GLOVES TO AVOID
BEING BURNED WHEN
TOUCHING THE VALVE.
At the end of the
program (CYCLE
COMPLETE), the door
will not open.
There is residual pressure
remaining in the sterilization
chamber at the end of the cycle.
NOTE: the display shows:
N O W LEVELLING
PLEASE WAIT…
Wait several minutes, until the pressure returns to 0.00
bar, and try to open the door again.
Check if the bacteriological filter is clogged and, if
necessary, replace it with a new one.
The procedure for storing the ambient temperature
(SET 0 bar function) was not executed correctly.
Contact the Technical Support Department (see
Appendix Z)
At the end of the cycle, the safety
door lock remains on.
Contact the Technical Support Department (see
Appendix Z).
There is water leaking
from underneath the
sterilizer.
Drain connectors or tubing (optional
external tank) not correctly
connected to the device.
Check the tightness of the fittings; if necessary,
reassemble, paying more attention to sealing.
Check that the tubes to the drain tank are completely
pushed onto the connectors; make sure that the plastic
ties have been applied.