11. TEST PROGRAMS
48. .
VACUUM TEST
To select the VACUUM TEST program, press the Test Selection key one or two times until
the display reads:
VACUUM TEST
- 0.80 bar
The Vacuum Test program is run with the sterilization chamber empty, except for the trays and
their supports.
NOTE
RUN THE VACUUM TEST AS THE FIRST CYCLE AFTER POWERING-ON THE EQUIPMENT.
To avoid the heating of the sterilization chamber influencing the variation of the vacuum value
measured during the Vacuum Test, the system is programmed to prevent running this test
when the temperature sensors of the sterilization chamber show a value higher than 5 C.
If you try to start the program with a higher temperature than indicated above, the display will
read:
WARNING!
PT1 OVERHEATING
- 0.80 bar
After a short time, the device will automatically return to STAND-B Y mode, ready for use.
NOTE
TO RAPIDLY LOWER THE TEMPERATURE OF THE CHAMBER LEAVE THE STERILIZER'S
DOOR OPEN UNTIL THE CORRECT TEMPERATURE IS REACHED.
Close the door and start the program with the S T AR T key.
NOTE
IN ADDITION, THE EQUIPMENT CHECKS THE PRINTER PAPER PRESENCE (OPTION) A N D, IF
A DATA RECORDER IS CONNECTED, THE PRESENCE OF THE FLASH CARD AND ITS
M E MO RY C AP A C I TY.
THE POSSIBLE WARNING MESSAGES, AND THE CONSEQUENT ACTIONS TO CARRY OUT,
ARE THE SAME AS DESCRIBED FOR A STANDARD STERILIZATION CYCLE.
The vacuum phase will begin immediately and the display reads:
VACUUM TEST
VACUUM PULSE
- 0.69 bar 01:02
The display shows the pressure (bar), and the total time from the start of the program.
When the pre-set pressure is reached (-0.80 bar) the pump stops and the pressure
stabilization phase begins (WAITING PERIOD). This lasts 5 minutes is shown on the display: