APPENDIX C M AI N TE N AN C E
65.
DANGER
BEFORE PERFORMING ORDINARY MAINTENANCE, MAKE SURE THAT
THE POWER SUPPLY CORD IS REMOVED FROM THE MAINS SOCKET.
WHEN IT IS NOT POSSIBLE, TURN OFF THE EXTERNAL BREAKER OF
THE EQUIPMENT POWER SUPPLY LINE.
IF THE EXTERNAL BREAKER IS F AR AW A Y O R, AT AN Y R AT E , NOT
VISIBLE TO THE MAINTAINANCE WORKER, PLACE A WORK IN
PROGRESS SIGN ON THE EXTERNAL BREAKER AFTER TURNING IT OFF.
MAINTENANCE
DESCRIPTION
Clean door gasket and
porthole (door plate)
Clean external surfaces
Clean sterilization
chamber and
accessories
Disinfect external
surfaces
To remove traces of lime, clean the door gasket of the container and the porthole (door plate)
with a clean, cotton cloth soaked in a weak solution of water and vinegar (or similar product).
Dry the surfaces and remove any residue before using the device.
Clean all the external parts using a clean cotton cloth dampened with water and, if needed, a
neutral detergent.
Dry the surfaces and remove any residue before using the device.
Clean the sterilization chamber, support and trays (and internal surfaces in general) with a
clean cotton cloth soaked in water and, if needed, use a small amount of neutral detergent.
Carefully rinse with distilled water, taking care not to leave any type of residue in the chamber
or on accessories.
NOTE
DO NOT USE SHARP OR POINTED INSTRUMENTS TO REMOVE LIME ENCRUSTATION FROM
THE STERILIZATION CHA MB ER . WHEN THERE ARE VISIBLE DEPOSITS, IMMEDIATELY
CHECK THE QUALITY OF THE DISTILLED WATER USED (SEE APPENDIX A,).
For the occasional disinfection of the external surfaces, you can use either denatured alcohol
or detergents with a small percentage of sodium hypochlorite (or equivalent).