9. RUNNING THE CYCLE
44. .
CHECKING THE
CYCLE DATA
REPORT (FOR
UNITS WITH
PRINTERS)
It is a good practice to check that the print report issued at the end of the sterilization program,
also specifies a positive outcome.
At the end of the cycle, the relevant data for the thermodynamic parameters of the sterilization,
i.e., temperature and pressure C and bar), and time (in minutes) of the sterilization cycle,
along with particular attention to the sterilization phase, will print automatically when the door
is opened.
Check the values on the print report and any additional indications for further confirmation of
sterilization.
The operator should sign in the space provided and file the document for possible future use.
If necessary, copies of the document can be used to identify the load (or parts of it) with the
date/time of sterilization and details of the type of cycle performed.
To select the number of copies to print, consult Chapter 6 - Configuration.
NOTE
THE OPERATOR CAN ALSO REQUEST AN EXTENDED PRINTOUT OF THE STERILIZATION
PROCESS DATA, INCLUDING THE RECORDED VALUES OF ALL THE SENSORS INSTALLED
ON THE MACHINE. TO START THIS PRINT FUNCTION, HOLD DOWN THE (ESC) KEY ON
THE CONTROL PANEL WHILE OPENING THE DOOR.
FOR COMPLETE DETAILS ABOUT PRINTING THE SUMMARY, PLEASE REFER TO THE
REPORT EXAMPLES SHOWN IN APPENDIX B, PROGRAMS.
MANUAL CYCLE
INTERRUPTION
The operator can manually interrupt the cycle at any time by pressing the START/STOP key
for three seconds. This command generates the error E999, because the cycle did not finish
correctly. Until it is safe to open the door, the unit will beep and the display will show:
MANUAL STOP
LEVELLING...
101.2 °C E999
- 0.47 bar 26:01
When safe conditions are reached, the machine activates a special procedure, first asking the
user to manually unlock the door by displaying the following instruction:
PRESS TO
UNLOCK THE DOOR
86.2°C E999
- 0.02 bar 26:01
Press the key to unlock the door.
The following message is then displayed:
MANUAL STOP
OPEN THE DOOR
85.8°C E999
- 0.01 bar 26:01
> 3s
Error code