APPENDIX E AL A RM S
73.
Alarm outside
the c y c l e
At the end of what has been described and having reached safe conditions, the machine
activates a special procedure, that asks the user to manually unlock the door:
PRESS TO
UNLOCK THE DOOR
101.0°C XXXX
0.01 bar 11:40
NOTE
THE ABOVE INDICATED MESSAGE IS SHOWN ONLY WHEN THE PRESSURE IN THE
CHAMBER IS WITHIN A SAFET LIMIT.
THE RELEASE OF THE LOCKING DEVICE IS NOT POSSIBLE WHEN THE PRESSURE VALUE
IS OUTSIDE THIS LIMIT.
Press the key to unlock the door lock mechanism; the following message will appear:
(Alarm Message)
OPEN THE DOOR
100.8°C XXXX
0.01 bar 11:42
Once the door is open, the user is asked to reset the system:
(Alarm Message)
RESET SYSTEM
95.5°C XXXX
0.00 bar 11:43
Perform a RESET (described below) and then turn-off the equipment and check the error or
make the repair.
NOTE
WHEN THE DOOR IS OPENED, THE REPORT (NORMAL OR EXTENDED DEPENDING ON THE
TYPE OF ALARM) WILL BE PRINTED FOR THE INTERRUPTED STERILIZATION P R OG R AM
AND THE ALARM THAT INTERVENED. CHECK THE DOCUMENT, INITIAL IT IN THE SPA CE
PROVIDED AND FILE IT IN A SUITABLE PLACE. REFER TO THE PRINT REPORT EXAMPLES
SHOWN IN APPENDIX B, PROGRAMS.
If the alarm intervenes outside the sterilization or test program, the display will show:
( Alarm Message)
ALARM!
80.5 °C XXXX
0.00 bar
Turn-off the equipment and check the alarm. Or, depending on the type of alarm:
( Alarm Message)
WARNING!
80.5 °C XXXX
0.00 bar
Alarm Message
Alarm Code
Alarm Message
Alarm Code