APPENDIX B – PROGRAMS
53.
APPENDIX B – P R O G R A M S
INTRODUCTION
The steam sterilizer is appropriate for almost all materials and instruments, so long as they are able to tolerate, without
damage, a minimum temperature of 121 °C.
The following material can typically be sterilized with steam:
– Stainless steel surgical/generic instruments;
– Carbon steel surgical/generic instruments;
– Rotating and/or vibrating instruments driven by compressed air (turbines) or mechanical transmission (counter-angles,
tooth scalers);
– Glass articles:
– Mineral-based articles;
– Articles made of heat-resistant plastic;
– Articles made of heat-resistant rubber;
– Heat-resistant textiles;
– Medical textiles (gauze, pads, etc.);
TO PREVENT THE INSTRUMENTS AND/OR MATERIALS FROM ELECTROLYTHIC CORROSION DURING THE
STERILIZATION PROCESS, PLEASE AV O ID DIRECT CONTACT BETWEEN THE FOLLOWING METALS:
ALUMINUM (AL) - NICKEL (NI);
CARBON STEEL – NICKEL (NI);
NICKEL (NI) – CHROME (CR);
COPPER (CU) – ALUMINUM (AL);
CARBON STEEL – COPPER (CU);
CHROME (CR) – COPPER (CU);
STAINLESS STEEL – ALUMINUM (AL);
CARBON STEEL – STAINLESS STEEL;
CHROME (CR) – STAINLESS STEEL.
ALWAYS SEPARATE THE INSTRUMENTS A ND /OR MATERIALS BY METAL TYPE AND ELECTROLYTHIC COMPATIBILITY.
DEPENDING ON THE CONFORMATION OF THE MATERIAL (SOLID, HOLLOW OR POROUS), ANY PACKAGING
(PAPER/PLASTIC ENVELOPE, STERILIZATION PAPER, CONTAINER, MUSLIN NAPKIN, ETC.) A N D ITS H E AT -
RESISTANCE, IT IS IMPORTANT THAT YOU CHOOSE THE APPROPRIATE PROGRAM BY REFERRING TO THE TABLE
SHOWN ON THE NEXT PAGE.
THE DEVICE MUST NOT BE USED FOR STERILIZING FLUIDS, LIQUIDS OR PHARMACEUTICAL
PRODUCTS.