APPENDIX C – M AI N TE N AN C E
67.
Replace bacteriological
filter
Replace the door gasket
When it is due to be changed, or when you notice visible clogging of the filter (when the filter
turns gray) unscrew the bacteriological filter from its support and replace it with a new one by
screwing it all the way down on the connector on the front of the machine.
A REPLACEMENT BACTERIOLOGICAL FILTER IS SUPPLIED WITH THE DEVICE. TO
REQUEST OTHERS, PLEASE REFER TO APPENDIX Z, TECHNICAL SUPPORT.
Check the inside of the door to ensure it is not hot and then remove the old gasket by hand.
Clean the door gasket seat to ensure it is debris free.
Install the new door gasket by pressing the gasket into its seat, first on top, then bottom, then
both sides. Once seated on 4 sides, continue to press the remaining gasket completely into its
seat