APPENDIX E – AL A RM S
74. .
which is automatically transformed to the message:
( Alarm Message)
RESET SYSTEM
80.5 °C XXXX
0.00 bar
Perform a RESET (described below) and then turn-off the device and check the alarm.
ALARMS THAT INTERVENE OUTSIDE OF A PROGRAM DO NOT PRODUCE A PRINTED
REPORT.
RESETTING THE
SYSTEM
Depending on the alarm, the system must be reset in one of two ways. (see the Alarm Code
List further below in this appendix):
1. Press and hold the PROGRAM SELECTION key for about 3 seconds.
A beep confirms the RESET.
NEVER TURN THE DEVICE OFF BEFORE TRYING TO EXECUTE A RESET
AS DESCRIBED ABOVE.
2. Turn-off the device and then power-on using the main switch.
Upon power-up, the sterilizer will perform its normal initial test.
After RESET, and any technical intervention necessary to eliminate the fault, the device will go
to STAND-BY mode, ready to execute a new program.