APPENDIX E – AL A RM S
82. .
Intervention of the steam generator
safety thermostat.
Manually rearm the thermostat(s) located on the back of the device (see
the Chapter 3 - Product Overview).
Unscrew the black plastic protection cap, press the white button until
you hear a click and replace the cap.
Turn-off (RESET) and then turn-on the device.
If the problem persists, Contact the Technical Support Department (see
Appendix Z).
Heating element safety thermostat
intervened.
Contact the Technical Support Department
(see Appendix Z).
Heating or steam generator heating
element malfunction.
Perform RESET following the instructions.
Check the quantity of the material in the sterilization chamber and make
sure that it does not exceed the maximum allowed quantity, depending
on the type of load.
(See the Summary Table in Appendix A, Technical Characteristics).
Steam leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with
water, and start the program again. If the gasket still leaks, replace the
gasket.
Problem in the plumbing circuit.
Contact the Technical Support Department
(see Appendix Z).
Problem in the plumbing circuit.
Pressure sensor (MPX) broken.
Contact the Technical Support Department
(see Appendix Z).
Incorrect connection of the
pressure sensor (MPX) to the
connector.
Pressure sensor (MPX) short
circuit.
Problem in the plumbing circuit.
Problem in the plumbing circuit.
Steam generator malfunction.
Problem in the plumbing circuit.
Steam generator malfunction.
Problem in the plumbing circuit.
Problem in the plumbing circuit.
Steam generator malfunction.
Problem in the plumbing circuit.
Steam generator malfunction.
Problem in the plumbing circuit.
Steam generator malfunction.
General operating problem.
General operating problem.
General operating problem.
General operating problem.