EasyManuals Logo

Shimadzu GC-2014 Operation Manual

Shimadzu GC-2014
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
4.7 Graphite ferrule
94
4.7 Graphite ferrule
If you are using packed column
Purpose of inspection
(1)
Checking for carrier gas leaks
If carrier gas leaks due to graphite deterioration, it causes poor reproducibility.
Check to see if graphite has deteriorated or not. If it has, replace with a new graphite ferrule.
(2)
Checking if graphite impurity is the cause of ghost peaks
Generally speaking, if ghost peaks appear in your analysis, there can be impurities in the column inlet
connection.
If your baseline drift is quite large even though using a sufficiently aged column, there can be impurities
in the column outlet connection.
In such cases, condition graphite ferrule by the following procedures.
Conditioning graphite ferrule
If the graphite impurity is the cause of ghost peaks, condition graphite ferrule by this procedure. (Perform
this just before you start analyzing. If you leave graphite ferrule for a while after conditioning, it can absorb
impurities again.)
Use gas burner. Put graphite ferrule into the blue flame, and heat it for 1 to 2 sec. after graphite turns red.
Indicates deterioration (no gap)
New graphite ferrule Graphite ferrule to be replaced
CAUTION
Hold graphite ferrule in such a way as to avoid burns.
Graphite ferrule

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals