4.11 FID
4
4.11 FID
4.11.1 Maintenance intervals
In the following cases, inspect / replace the igniter, and also inspect / maintain the FID as required.
When you can not ignite the FID
When peaks do not appear
When noise is significant
When peak shape is abnormal
WARNING
Before inspecting the FID, ALWAYS make sure the following are observed.
1. Stop supplying hydrogen gas to extinguish the hydrogen flame.
2. Let the detector cool until its temperature drops below 40C.
3. Turn off the instrument.
4. Remove capillary column from the FID.
RISK OF BURNSNEVER inspect / maintenance FID, when it is at high temperature.
CAUTION
To prevent screws and nuts from burning, DO NOT turn them or force the fitted part to move when the FID is at
high temperature.
NOTE
For the handling of FID, refer to the instruction manual Chp.13 [Detectors "DET"].

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals