V
Safety Precautions
Gas Usage Precautions
The following information explains important points concerning the handling of high-pressure gas. Be sure to
read all information carefully and follow all instructions.
For information on gas types, purity, and supply pressures, refer to the GC Instruction Manual [2.6 Gas supply
plumbing].
When handling high-pressure gas cylinders:
Be sure to install gas cylinders in a well-ventilated outdoor area, away from direct sunlight. Use tubing to
bring the gases to the lab.
DO NOT allow the temperature of the gas cylinders to exceed 40°C. Keep all flammable items at least
2 m away from gas cylinders.
Be sure the work area is well ventilated when using high-pressure gases. Perform a daily leak check
before starting operation using soapy water, etc.
When using flammable gases (e.g. hydrogen), DO NOT smoke or use fire within 5 m of the instrument.
Locate fire extinguishers nearby for use in case of accident.
Be sure to secure gas cylinders in place with cylinder clamps or ropes to prevent them from falling over.
Be sure to use only oil-free reducing valves. DO NOT use pipes that have oil on the inner surface.
Be sure to close the main valve of a gas cylinder immediately after using that gas.
Be sure to observe all laws and regulations in your country.
WARNING
WHEN USING HYDROGEN GAS
Risk of explosion.
If hydrogen accumulates in the column oven, it could cause an explosion.
DO NOT leave the hydrogen pressure control valve open when the system is not in use. When not in use, turn
the valve OFF and cover the column mounting port with a sealing plug.
Capping the column mounting port of detectors
Packed column mounting port Capillary column mounting port
Sealing plug
Capillary adaptor
Detector
Graphite ferrule with a wire
Column nut (Split nut)

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals