63
3.3 Analysis using capillary column
3
Checking gas leak
After installation of column, etc., has been completed, perform a leak check. If there is leakage in the flow line,
you will not obtain high reproducibility in your analysis, and also can waste carrier gas.
Specify temperature, etc., temporarily when checking.
1
Prepare carrier gas.
2
Turn on the power.
3
Press .
Specify "Start time" to the required time for the leak
check.
Verify that "Yes" is selected on "Start Temp/Det".
OFF ON