1.3 Specifications
10
1.3 Specifications
Column oven
Temperature program
Injection port
Item Specification
Dimensions (mm) 250 (W) 360 (H) 175 (D)
Volume (L) 15.8
Temperature range Room temperature + 10C to 400C
50C to 400C (When using liquid carbon dioxide gas.)
Temperature accuracy ±1% (K) (Calibration possible at 0.01C)
Temperature deviation Within 2C (on a 200 mm diameter column holder)
Temperature stability Within ±0.05C
Temperature coefficient 0.01C/C
Range of linear temperature increase 115 V model
30C/min up to 150C
20C/min up to 250C
10C/min up to 380C
7C/min up to 400C
230 V model
60C/min up to 150C
40C/min up to 250C
20C/min up to 380C
15C/min up to 400C
Cooling speed To cool from 300C to 50C, approx. 6 minutes.
Overheat protection Programmable up to 420C
(Also, you have another protection separately: fixed at 500C.)
Item Specification
Program ramps 20 ramps in total (Heating and cooling available)
Setting 0.1C increments
Program rate 250 to 250C/min, 0.01C/min increments
All program total time Maximum 9999.99 minutes
Item Specification
Temperature range Up to 400C
Temperature setting 1C increments
Overheat protection Programmable up to 420C
Injection unit Split/Splitless injection, Packed injection (direct)

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)