5.1 Trouble shooting
108
5.1 Trouble shooting
If you find the following problems, perform inspection/maintenance.
Problem Possible cause Solution Reference
Poor retention time
reproducibility
Septum Condition septum. Refer to "4.4
Septum" P. 87.
The position of silica wool in
glass insert
Re-pack silica wool. Refer to "4.6 Glass
insert" P. 91.
Stains in glass insert Cleanse glass insert. Refer to "4.6 Glass
insert" P. 91.
Poor peak area
reproducibility
Septum Condition septum. Refer to "4.4
Septum" P. 87.
The position of silica wool in
glass insert
Re-pack silica wool. Refer to "4.6 Glass
insert" P. 91.
Stains in glass insert Cleanse glass insert. Refer to "4.6 Glass
insert" P. 91.
Ghost peaks are
detected.
Septum Condition septum. Refer to "4.4
Septum" P. 87.
The position of silica wool in
glass insert
Re-pack silica wool. Refer to "4.6 Glass
insert" P. 91.
Stains in glass insert Cleanse glass insert. Refer to "4.6 Glass
insert" P. 91.
Capillary column Condition capillary column. Refer to "4.9
Capillary column" P.
97.
Baseline is unstable. Capillary column Condition capillary column. Refer to "4.9
Capillary column" P.
97.
Ghost peaks are detected
when temp. is rising.
Graphite ferrule Inspect graphite ferrule. Refer to "4.7
Graphite ferrule" P.
94.
Baseline drifts greatly
when temp. is rising.
Graphite ferrule Inspect graphite ferrule. Refer to "4.7
Graphite ferrule" P.
94.
Carrier gas is leaking. O-ring for glass insert Tighten to fix it. Refer to "4.5 O-ring
for glass insert" P.
89.
Carrier gas still leaks
even though you have
tightened the O-ring.
O-ring for glass insert Replace it since it may be
deteriorated.
Refer to "4.5 O-ring
for glass insert" P.
89.

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals