5.4 Key function list
126
5.4 Key function list
Key PF menu Description
Start GC, Stop GC, File, Clean Up, Start Seq,
Stop Seq, Maint INJ, Anal
• Use to start/stop the GC. On sub screen, you can specify
file management, clean up, injection port maintenance,
etc.
File, Line Config, Customiz, Print, Next line • Use when specifying frequently used parameters. On
[Line Config] screen, you can specify the combination of
injection port, detectors, etc.
Temp Mon, Flow Mon, Zero Adj, Zero Free, Up,
Down, Chng Graph, Chng Line
• Use to monitor the GC status and the analyzing process.
Adding to the each component temp., pressure, flow rate,
etc., chromatograms are also displayed.
Del Line, Ins Line, Fan Off, Fan On, Print • Specifies oven temperatures and oven temperature
programs.
Print, Next Inj
• Specifies injection port temperature.
Column, Gas Saver, On/Off, Press Prog, Flow
Prog, Split Prog, Purge, Advanced, Offset, Next
Flow
• Specifies parameters of carrier gas flow rate such as
pressure, flow rate, split ratio.
Ignite, Det gas, Ign. Set, Next Det • Specifies detector temp., detector gas (when APC is
installed), output signal, etc.
AOC parameter • Specifies the auto injector, AOC-20i, and the auto
sampler, AOC-20s.
AUX temp. setting • Specifies temperature controls other than those for
column oven, injection ports, and detectors.
AUX APC • Specifies pressure, flow rate, type of gas, etc., of AUX
APC.
AUX AMC • Specifies flow rate, type of gas, etc., of AUX AMC.
CRG • Specifies the setting of CRG (Cryogenic work attachment:
optional).
1.Standard Diagnosis • Use to diagnose each component of the GC. Performing
standard diagnosis periodically helps the GC
management, and is very important to avoid any
malfunction that may occur.
2. Log Reading
Menu
1. GC Operation Log • Displays the log of the power and system On/Off.
2. Analysis Log • Displays the log to see if analyses have done completely,
or the controlled values have been in line with the
specified values.
3. Parameter Log • Displays the log of key operation and parameter log
changing.
4. Error Log • Displays the log of displayed error messages.
5. Diagnostic Log • Displays the log of diagnosis.
3. Analysis Counter • You can display the time for replacement of septa and
glass inserts as warning message.
4. Coolant Consumption Counter • Integrate the time for which the CRG (optional) has been
ON.
You can set a warning message to appear when the
integrated time exceeds your specified time.