VII
Safety Precautions
Installation Location
To ensure safe operation, contact your Shimadzu representative if product installation, adjustment, or re-
installation (after the product is moved) is required.
Clearance on Sides and Rear Face
Refer to the instruction manual for information about power supply, wiring, and piping.
WARNING
FIRE
Be sure to keep flammable materials away from exhaust opening, since exhaust air is extremely hot.
CAUTION
When the column oven is cooled, extremely hot
air is exhausted from the rear face exhaust vent.
As a result, when installing the instrument be
sure to observe the following.
1. DO NOT place flammable materials behind
the instrument.
2. Be sure to maintain a clearance of 400 mm or
more between the rear cover and wall.
3. Be sure to maintain a clearance of 50 mm or
more on the left side of the instrument.
4. To open/close the oven door, be sure to
maintain a clearance of 100 mm or more on
the right side of the instrument.
5. Be sure to maintain sufficient space in the rear
for maintenance and inspection of the rear face of the instrument.
Minimum 50 mm
Minimum
400 mm
Depth: 607 mm
Minimum 100 mm
Height: 690 mm
Width: 400 mm
TOP VIEW
FRONT VIEW

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals