5.2 Error messages (extract)
124
Use time (Replacement time)
Fan:61320 hours (based on 8 hours per day:approx. 21 years, 24 hours: approx. 7 years)
Sensor: 26280 hours (based on 8 hours per day:approx. 9 years, 24 hours: approx. 3 years; at 300 C)
NOTE
The replacement time shows the number of years at which a replacement is recommended. Note that this is not
the guaranteed service life of each part.
If a warning massage is displayed for the sensor of a unit, consider that the other units that you started to use at
the same time are also close to the replacement time.
It is recommended that these parts should be replaced at the same time even if their warning messages are not
displayed.

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz