3.2 Analysis using packed column
42
3
Specify injection port temperature.
(1) Press .
(2) Specify the temperature.
4
Specify column oven temperature.
(1) Press .
(2) Specify column initial temperature, and
parameters of temperature program.
NOTE
Normally, specify the temperature of injection port and
detectors at approx. 20C higher than column oven
temperature (or final temperature when executing
temperature program.)
CAUTION
DO NOT specify higher than the allowable
temperature limit of the packing material.
NOTE
To avoid detector contamination, it is highly
recommended to not specify/make column oven
temperature higher than the detector temperature.

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals