3.2 Analysis using packed column
3
When the temperature and flow rate of each
component are stabilized at their specified values,
[STATUS] indicator illuminates green, and the
preparation has been completed.
At this time, by referring to the detector polarity
setting, the GC displays "Monitor injection screen"
that indicates which injection port (left or right) you
should inject.
When the system is "Ready", this screen keeps
appearing automatically after 1 minute of no key
operation, even though you move to the other
screens.
When using the GC in single flow line
mode.
At the procedure [1. Specify the dual AFC setting.],
select "Single L" or "Single R" on [Control mode], and
specify the flow rate.
Here, remove the column in which carrier gas is not
flowed.
At the procedure [2. Specify the detectors.], select
"Single L" or "Single R" on [Control mode].
[Signal polarity] is not displayed since its setting is not
required.
The monitor injection screen in "single mode"
indicates which injection port (left or right) you
should inject by referring the control mode
setting (L or R) of dual AFC.
Since [Zero to ready] has been set to [On] in default,
immediately after [STATUS] indicator illuminates green,
the detector signal is stabilized at approx. 0 V.