45
3.2 Analysis using packed column
3
When the temperature and flow rate of each
component are stabilized at their specified values,
[STATUS] indicator illuminates green, and the
preparation has been completed.
At this time, by referring to the detector polarity
setting, the GC displays "Monitor injection screen"
that indicates which injection port (left or right) you
should inject.
When the system is "Ready", this screen keeps
appearing automatically after 1 minute of no key
operation, even though you move to the other
screens.
When using the GC in single flow line
mode.
At the procedure [1. Specify the dual AFC setting.],
select "Single L" or "Single R" on [Control mode], and
specify the flow rate.
Here, remove the column in which carrier gas is not
flowed.
At the procedure [2. Specify the detectors.], select
"Single L" or "Single R" on [Control mode].
[Signal polarity] is not displayed since its setting is not
required.
The monitor injection screen in "single mode"
indicates which injection port (left or right) you
should inject by referring the control mode
setting (L or R) of dual AFC.
NOTE
Since [Zero to ready] has been set to [On] in default,
immediately after [STATUS] indicator illuminates green,
the detector signal is stabilized at approx. 0 V.

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)