Shimadzu LC-10/20 System 4 Using the control module
Rinse Mode
Sets the time when the rinsing will be performed. Available options are No
Rinsing , Before Aspiration , After Aspiration and Before and After
Aspiration.
Cooler Temperature
Enables the cooling of the autosampler. This option is available only after
the Sample Cooler Installed option was checked in the Shimadzu LC-
10/20 System Setup - Autosampler dialog. The field than allows to set the
desired temperature.
Check Cooler Temperature
If checked, doesn't allow the Instrument to become READY until the
temperature is not close to the temperature set in the Cooler
Temperature field (with the tolerance of 1,5 °C).
Control Vial Needle Stroke
Sets the needle depth used when aspirating from the Control Rack (used
with the SIL-20 samplers).
Rinse Pump
Section enabled by checking the Rinse Pump Installed option in the
Shimadzu LC-10/20 System Setup - Autosampler dialog.
Rinse Method
Sets the method for the Rinse Pump . Port Only, Pump Only, Pump
Then Port, Port Then Pump, Pump And Port Between Analyses and
Auxiliary 1 (2) methods are available.
Rinse Time
Sets the length of the rinse. Possible values are integers in the range
1 - 9.
Rack Changer
Section enabled by checking the Rack Changer Installed option in the
Shimadzu LC-10/20 System Setup - Autosampler dialog.
Rack Type D..A
Sets the type of the rack for each position in the rack changer.
Needle Stroke D..A
Sets the needle stroke depth (in mm) for each particular rack in the
rack changer.
Note: When the Rack Changer tray type has been selected, the common
Needle Stroke parameter is not used and thus is disabled. Moreover, do
not use values less than 20 for rack types 4, 5, 14 or 15.
- 48 -