38
8017171/115Z/V3-0/2019-01| SICKO P E R A T I N G I N S T R U C T I O N S | VISIC100SF
Subject to change without notice
INSTALLATION
12. Close the device:
Tip measuring unit up.
Screw the four screws with the SW4 Allen key.
Position the enclosure cover on the front side of the device.
Screw the two screws on the enclosure cover tight with the SW4 Allen key.
13. Visual check: Status LED should be green. The following reasons can cause the Status LED not
to be green:
LED switch on circuit board switched off. (Factory setting: LED switch is set to “On”) Switch
Figure, see “LED switch position on the circuit board”, page 29.
Enclosure cover not fitted (Status LED red).
Gas sensors in heating up phase (Status LED red for max. 30 minutes).
Check the plug on the circuit board when the Status LED is not on.
Active maintenance and malfunction states. To retrieve the maintenance request and
malfunction messages, see “Testing the analog output for the K-value with submenu item
“AO1, page 60
. Malfunction and maintenance request Code Tables, see “Device error
coding”, page 95
and see “Description of maintenance requests”, page 96.