ELECTRICAL INSTALLATION 5
77
8026362 / V1-0/2022-03|SICK
Subject to change without notice
SUPPLEMENTARY OPERATING INSTRUCTIONS | VMS4100/5100
5.2 Connecting the voltage supply
Fig. 74: Pin assignment of the -X100 terminal block in the cabinet
Terminal Color Signal Function
-X100/1.1 Gray L Line voltage 100…264 V AC / 50…60 Hz
(phase)
-X100/1.4 Blue N Line voltage 100 ... 264 V AC / 50 ... 60 Hz
(neutral conductor)
-X100/1.6 Green-
yellow
PE Protective conductor
Tab. 9: Pin assignment of the -X100 terminal block in the cabinet
NOTE!
Make sure that cables are securely connected.
No visible metal surfaces are permitted on the wires.
Tighten the coupling nuts to provide strain relief on the cabinet. Doing so also
maintains the enclosure rating.