DCA Vantage Operator’s Guide 137
Troubleshooting
E55
Ethernet port
communication
failure
An attempt to
communicate via the
Ethernet serial port
failed.
1. Verify the cables connecting
the DCA Vantage system and
the computer interface are
securely connected.
2. Verify the DCA Vantage
system has the correct
configuration for the host
device (baud rate, parity).
3. Disable and enable the
Ethernet port.
4. Restart the system.
E56
Ethernet port
communication
error
Communication with the
Ethernet port is not
successful, even after
numerous attempts. This
error occurs after an E55
error.
1. Verify the cables connecting
the DCA Vantage system and
the computer interface are
securely connected.
2. Verify the DCA Vantage
system has the correct
configuration for the host
device (baud rate, parity).
3. Disable and enable the
Ethernet port.
4. Restart the system.
E60
Database
Failure
The database is not
functioning properly. The
database may be
corrupted or the system
was not properly
powered down.
1. Turn off the instrument from
the Home screen and wait until
system shutdown is complete.
2. Restart the system.
E61
Database read
error
When attempting to read
the database, the read
operation fails, resulting
in the return of no data
or corrupt data.
1. Turn off the instrument from
the Home screen and wait until
system shutdown is complete.
2. Restart the system.
E62
Database write
error
An error is returned from
a database write
operation. The database
may be corrupted.
Restart the system.
Error Code Description Corrective Action

Table of Contents