Functions
7.1 Selecting Flow Units
FUE1010 IP65 NEMA 4X
Operating Instructions, 12/2014, A5E03086491-AC
99
Note
NETFLOW (default) is best for applications where there may be zero flow for long periods. It
minimizes false Totalizer register increments due to data scatter. Press the <Down Arrow> to
accept the default setting.
Selecting Totalizer modes
1. Press the <MENU> key and [Meter Type] will be highlighted.
2. Press the <Right Arrow> to [Dual Path Flow] and then press <ENTER>.
3. The [Dual Path Flow] menu with appear with [Chan/Path Setup] menu item highlighted.
4. Scroll down to the [Totalizer Mode] menu and press the <Right Arrow> to select the
Totalizer Mode option list.
5. Press the <Up/Down Arrows> to select the desired mode.
6. Press <ENTER> to store selection.
Totalizer Mode Controls
From the RS-232 serial port all of the Totalizer commands listed below can be executed
using PC keyboard function keys via VT100 terminal key emulation.
Note
Si
-Ware or HyperTerminal should be in the Data Display mode when invoking the function
keys referenced in the table below. Use the key sequence <Ctrl L> to display the Data
Display mode.