Functions
7.1 Selecting Flow Units
FUE1010 IP65 NEMA 4X
100 Operating Instructions, 12/2014, A5E03086491-AC
Connect the flow meter to your PC. Refer to Appendix A for communications setup
procedures, if needed.
1. Access Si-Ware or, if using a PC, access HyperTerminal from the PC [Programs] menu,
then select [HyperTerminal].
2. In [Connection Description] dialog box, enter a connection name (e.g. FUE1010). Click
[OK].
3. In [Phone Number] dialog box, select [Direct to COM 1 (or COM 2)]. Click [OK] to select.
4. In [Properties] dialog box, enter RS-232 parameters. Click [OK].
5. At terminal screen, click [File]. Select [Properties].
6. Select [Settings] tab. At [Emulation] box, select [VT-100].
7. Select [ASCII Setup]. In [ASCII Sending] uncheck boxes. In [ASCII Receiving] check
[Append line feeds to incoming line ends.]. Click [OK].
8. At the Terminal screen, press <ENTER> and the Data Display mode appears.
9. If not, to enter the Data Display mode type MENU and then press <Ctrl L>.