7.10 Analog Output Trim
FUE1010 IP65 NEMA 4X
Operating Instructions, 12/2014, A5E03086491-AC
125
To calculate a voltage output:
1. Set up the multi-meter to read volts, then connect it to the supply and return terminals of
the voltage output under test.
2. Move the highlight to the port to be tested by pressing <Up/Down Arrow>, press <Right
Arrow>, then press <Down Arrow> to move the cursor to [Trim @ 2V].
3. Press <ENTER>. This triggers a 2.00 Volts pop-up window. The multi-meter will now be
reading 2.00 Volts.
4. If the multi-meter reading does not match, use the numeric keys to type in the multi-meter
reading.
5. Press <ENTER> to register setting. This adjusts the flow meter’s DAC (digital-to-analog
converter) so that a 2.00 Volts output corresponds with 2.00 Volts on the multi-meter.
6. Re-check the multi-meter to make sure that it is now reading 2.00 Volts.
Pgen Output Trim
Note
Can be trimmed to within 1.25 Hz of nominal.
To calibrate a pulse rate output
1. Connect a frequency counter to the signal and reference terminals of the pulse rate
output under test.
2. Move the highlight to the port to be tested by pressing <Up/Down Arrow>, press <Right
Arrow>, then press the <Down Arrow> to move the cursor to Trim @ 1kHz.
3. Press <ENTER>. This triggers a 1 kHz pop-up window. The frequency counter will now
be reading 1 kHz.
4. If the frequency counter reading does not match, use the numeric keys to type in the
frequency counter reading.