Commissioning
6.2 Empty Pipe Set
FUE1010 IP65 NEMA 4X
90 Operating Instructions, 12/2014, A5E03086491-AC
3. Press the <Down Arrow> to [Actual MTY] then press <ENTER>.
Empty Pipe Press [ENT]
appears on the menu prompt line.
4. Empty the pipe completely, then press <ENTER>.
Fill Pipe Press [ENT]
appears on the menu prompt line.
5. Refill the pipe completely, then press <ENTER>.
Using the MTYmatic command
You can repeat MTYmatic (performed during the Initial Makeup) to correct an inaccurate
Actual MTY setting if conditions do not allow you to repeat the Actual Empty procedure.
Note
IMPORTANT
Only use the MTYmatic procedu
re when the pipe is full.