6.11 Activation of write protection
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02
41
6.11
Activation of write protection
In the case of an officially verified SIWAREX WP231, write protection must be provided to
prevent the changing of data relevant to the calibration. Insert a wire jumper between the
module terminals P and PR for this purpose.
Table 6- 8 Activation of write protection
Labeling
Function
P P terminal for activation of write protection
PR PR terminal for activation of write protection