Messages
9.5 Message list
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02
135
System status message list
Operating errors (code 1000 to 1999)
sorted by code no.
1000 operating error exists 1000 Group message, at least one operating error exists.
1001 Watchdog 1001 Watchdog, error is displayed for at least 10 seconds. A serious error
has occurred in the function of SIWAREX, e.g. program error, severe
electromagnetic influence on device, etc.
Contact the SIWAREX Support if the error occurs multiple times.
1002 RAM error 1002 RAM error. An error has occurred in the memory; the memory con-
tent is no longer correct. The module must be switched off. If the
error occurs again, SIWAREX is defective.
1003 Checksum incorrect parameter 1003 Checksum error at parameter. Critical error because the parameters
are no longer safe.
1004 Checksum incorrect program 1004 Checksum error program code. Critical error because the program is
no longer safe.
1006 logbook error 1006 Error when writing/deleting,or logbook full
1102 ADU error 1102 AD converter error when reading in the measured value. If the error
occurs again, make sure that the EMC recommendations are ob-
served (chapter EMC-compliant setup (Page 25)).
1104 Undervoltage 1104 Undervoltage at sensor cables
1105 Overload 1105 Overload of scale (ca. 110%)
1106 Underload 1106 Underload of scale (ca. -10%)
1107 Legal trade display failure 1107 The SecureDisplay legal trade display no longer communicates with
the module
Technology error message list
Technology error (code 2000 to 4999)
2000 Technology error exists 2000 Group message, at least one technology error exists
2001 Timeout tare/set to zero 2001 Taring of scale or set to zero is not possible because a standstill was not
reached during the standstill time. The command was discarded.
2002 Trace overload 2002 The set recording rate for trace function cannot be processed. Set a slow-
er recording rate (chapter "Trace recording cycle (Page 95)")
2003 Set to zero not possible 2003 The switch-on weight is outside the configured value range for set to zero
in data record DR 3 defined by the maximum positive and negative weight.
Data and operating errors message list
Data and operating errors
(code 5000 to 8999)
5000 Data and operating error exists 5000 Group error, a bit is set in the data and operating error bits
5001 Command code or data record unknown 5001 Command code or data record is not known with current application
5002 Command or data change not possible
because write protection is active
5002 The command or changing of data is not possible due to write protec-
tion. The data record was rejected.