SPECIFICATIONS
19
11. Software calibration control.
The 6330 is calibrated through the front panel keypad. All calibration information is
stored in non-volatile memory. This allows the instrument to be completely calibrated
without removing any covers and exposing the technician to hazardous voltages.
12. Current monitoring.
The LCD display allows monitoring of current down to 10 microamps AC and 10
microamps DC. This allows the 6330 to be used even when test requirements only
allow a very low level of acceptable leakage current.
13. Line and load regulation.
This system maintains the output voltage to within 1% from no load to full load and
over the line voltage range to ensure that test results remain consistent and within
safety agency requirements.
14. PLC remote inputs and outputs.
The standard 9- pin interfaces provide outputs for Pass, Fail, and Test in Process.
Inputs include Test, Reset, Interlock, and Memory Recall. This gives the user all the
basic remotes required to configure the instrument through simple PLC relay control.
15. Flashing high voltage indicator.
A flashing LED located directly over the high voltage terminal clearly indicates when
high voltage is active to provide maximum operator safety.
16. Single Step Test Mode.
In this mode, the 6330 can be configured to perform each test individually as a single
step. This provides the manufacturers flexibility in configuring their testing
procedure.
17. Insulation Resistance and Ground Bond Test Modes.
This allows the operator to perform an Insulation Resistance or Ground Bond tests
with the Dielectric Withstand (Hipot) test, thus eliminating the need for a separate
pieces of test equipment when IR and Ground Bond testing is required.
18. Continuous Duty Cycle Ratings.
An advantage over other instruments in its class, the 6330 does not have any
requirement to be shut down during the day even when used at full output.
19. Milliohm Offset Capability.
This feature minimizes the effect of test lead or test fixture resistance.
20. Adjustable Output Current and Milliohm Trip Currents.
The 6330 is versatile enough to meet a wide variety of safety agency specifications.
21. Key Lockout.
When enabled, Key Lockout disables all front panel keys except TEST and RESET.
22. Memory Lock

Table of Contents